Empirical formula for the HPGe-detector efficiency dependence on energy and distance
The full-energy-peak efficiency calibration for a HPGe detector has been constructed in the γ -ray energy range from E γ = 50–1500 keV, using 8 standard point sources of gamma radiation, namely Na-22, Co-57, Co-60, Cd-109, Ba-133, Cs-137, Eu-152, Am-241. The measuring system equipped with sample ho...