Empirical formula for the HPGe-detector efficiency dependence on energy and distance

The full-energy-peak efficiency calibration for a HPGe detector has been constructed in the γ -ray energy range from E γ  = 50–1500 keV, using 8 standard point sources of gamma radiation, namely Na-22, Co-57, Co-60, Cd-109, Ba-133, Cs-137, Eu-152, Am-241. The measuring system equipped with sample ho...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radioanalytical and nuclear chemistry 2019-03, Vol.319 (3), p.1315-1319
Hauptverfasser: Pylypchynets, I., Lengyel, A., Parlag, O., Maslyuk, V., Potoki, I.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The full-energy-peak efficiency calibration for a HPGe detector has been constructed in the γ -ray energy range from E γ  = 50–1500 keV, using 8 standard point sources of gamma radiation, namely Na-22, Co-57, Co-60, Cd-109, Ba-133, Cs-137, Eu-152, Am-241. The measuring system equipped with sample holder to investigate the source-to-detector distance ( x ) dependence for 5–24.5 cm range. The universal empirical formula was proposed to calculate the detector efficiency depending on energy E γ and distance ( x ). The formula was tested for several measurements and broadened energy rage up to 3 MeV.