Combination of sorption tube sampling and thermal desorption with hollow waveguide FT-IR spectroscopy for atmospheric trace gas analysis: Determination of

The determination of organic trace gases in the ambient environment at the lower ppb level is demonstrated based on a novel technique combining sorption tube sampling on Molsieve and Carbosieve S-III, thermal desorption, and detection of the trace analyte by hollow waveguide Fourier transform infrar...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2004-01, Vol.76 (2), p.464
Hauptverfasser: Pogodina, Olga A, Pustogov, Victor V, de Melas, Fabiano, Haberhauer-Troyer, Christina
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The determination of organic trace gases in the ambient environment at the lower ppb level is demonstrated based on a novel technique combining sorption tube sampling on Molsieve and Carbosieve S-III, thermal desorption, and detection of the trace analyte by hollow waveguide Fourier transform infrared (HWG-FT-IR) spectroscopy. While ethene concentrations of ~5 ppm can be directly observed using HWG-FT-IR, enrichment factors of up to 5000 were achieved by sorption tube sampling and thermal desorption. Detection limits of ~1 ppb are reported. Efficient enrichment by the sampling tube is achieved due to the favorable internal volume (~0.4 cm3 at a length of 470 mm) of the hollow waveguide serving as a miniaturized gas cell. This new method was validated for ethene by thermodesorption-cryofocusing-GC-FID as the reference method. Analytical performance has been compared for standard gas mixtures and for ethene measurements in urban air. Finally, ethene data from a sampling campaign at two alpine sites in Tyrol/Austria are presented. [PUBLICATION ABSTRACT]
ISSN:0003-2700
1520-6882