Portable Gas Chromatograph with Tunable Retention and Sensor Array Detection for Determination of Complex Vapor Mixtures

A prototype portable gas chromatograph that combines a multiadsorbent preconcentrator/focuser, a tandem-column separation stage with individual column temperature control and junction point pressure modulation, and a detector consisting of an integrated array of polymer-coated surface acoustic wave...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2003-03, Vol.75 (6), p.1400-1409
Hauptverfasser: Lu, Chia-Jung, Whiting, Joshua, Sacks, Richard D, Zellers, Edward T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A prototype portable gas chromatograph that combines a multiadsorbent preconcentrator/focuser, a tandem-column separation stage with individual column temperature control and junction point pressure modulation, and a detector consisting of an integrated array of polymer-coated surface acoustic wave microsensors is described. Using scheduled first-column stop-flow intervals and independent temperature programming of the two columns, it is possible to adjust the retention of eluting analyte vapors to maximize vapor recognition with the microsensor array and minimize the time of analysis. A retention window approach is combined with Monte Carlo simulations to guide retention tuning requirements and facilitate pattern recognition analyses. The determination of a 30-vapor mixture of common indoor air contaminants in
ISSN:0003-2700
1520-6882
DOI:10.1021/ac026092n