A Planar Amperometric Creatinine Biosensor Employing an Insoluble Oxidizing Agent for Removing Redox-Active Interferences

A planar microchip-based creatinine biosensor employing an oxidizing layer (e.g., a PbO2 film), where interfering redox-active substances are broken (i.e., oxidized) to redox-inactive products, was developed to facilitate the microfabrication of the sensor and to provide improved, reliable determina...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2001-12, Vol.73 (24), p.5965-5971
Hauptverfasser: Shin, Jae Ho, Choi, Yong Suk, Lee, Han Jin, Choi, Sung Hyuk, Ha, Jeonghan, Yoon, In Jun, Nam, Hakhyun, Cha, Geun Sig
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A planar microchip-based creatinine biosensor employing an oxidizing layer (e.g., a PbO2 film), where interfering redox-active substances are broken (i.e., oxidized) to redox-inactive products, was developed to facilitate the microfabrication of the sensor and to provide improved, reliable determination of creatinine in physiological samples. The feasibility of using hydrophilic polyurethanes in permselective barrier membranes for creatinine biosensors and the effect of adding a silanizing agent (adhesion promoter) on the sensor performance (e.g., sensitivity, stability, and lifetime) are described. The proposed creatinine microsensor with a three-layer configuration, i.e., enzyme, protecting, and oxidizing layers, exhibits good electrochemical performance in terms of response time (t 95% = 98 s at 100 → 200 μM creatinine change), linearity (1−1000 μM, r = 0.9997), detection limit (0.8 μM), and lifetime (∼35 days). The creatinine biosensor devised in a differential sensing arrangement that compensates the erroneous results from creatine is considered to be suitable for assay of serum specimens.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac010497a