A Device for Thermal Treatment of Luminescence Ionizing-Radiation Detectors Intended for Radiation Nondestructive Testing
The structure and principle of operation of a heating unit intended for the thermal and radiation-thermal treatment of luminescence ionizing-radiation detectors aimed at modifying their dosimetric characteristics are described. The unit provides for linear heating of detectors in the range of 50–700...
Gespeichert in:
Veröffentlicht in: | Russian journal of nondestructive testing 2018-11, Vol.54 (11), p.792-796 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The structure and principle of operation of a heating unit intended for the thermal and radiation-thermal treatment of luminescence ionizing-radiation detectors aimed at modifying their dosimetric characteristics are described. The unit provides for linear heating of detectors in the range of 50–700°C at a given rate from 1 to 15°C/s, as well as for isothermal exposure of detectors. A description of unit operation is given as well as the results of trial tests that prove a high accuracy in reproducing the temperature and heating rate, acceptable for problems being solved. |
---|---|
ISSN: | 1061-8309 1608-3385 |
DOI: | 10.1134/S1061830918110062 |