An equivalent input noise level criterion for hearing aids
The purpose of this project was to establish a maximum acceptable equivalent input noise level (EINL) for hearing aids. It was found that, if the version of the National Acoustic Laboratories (NAL) procedure for selecting the gain and frequency response of hearing aids, which includes modifications...
Gespeichert in:
Veröffentlicht in: | Bulletin of prosthetics research 1996-10, Vol.33 (4), p.355-362 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this project was to establish a maximum acceptable equivalent input noise level (EINL) for hearing aids. It was found that, if the version of the National Acoustic Laboratories (NAL) procedure for selecting the gain and frequency response of hearing aids, which includes modifications for severe and profound hearing loss is used, the EINL criterion can be relaxed for higher gain hearing aids, as a function of the gain of the aid. The relationship between relaxation of the criterion and 2-cc coupler gain is nonlinear, in different amounts, at the various frequencies, so no simple rule can be used to describe the manner in which the criterion can be relaxed. |
---|---|
ISSN: | 0748-7711 0007-506X 1938-1352 |