A comparison of INAA and ICP-MS/ICP-AES methods for the analysis of meteorite samples

Instrumental neutron activation analysis (INAA), inductively coupled plasma atomic emission spectrometry (ICP-AES) and ICP mass spectrometry (ICP-MS) (hereafter, ICPs) were applied to meteorite samples for the determination of elemental content. The analytical applicability and suitability of the th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radioanalytical and nuclear chemistry 2018-12, Vol.318 (3), p.1681-1687
Hauptverfasser: Ebihara, Mitsuru, Shirai, Naoki, Bennett, John W., Stopic, Attila J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Instrumental neutron activation analysis (INAA), inductively coupled plasma atomic emission spectrometry (ICP-AES) and ICP mass spectrometry (ICP-MS) (hereafter, ICPs) were applied to meteorite samples for the determination of elemental content. The analytical applicability and suitability of the three methods have been compared. Those comparisons led to the refinement of our analytical procedures for INAA and ICPs, yielding more reliable data. Our INAA data proved to be reliable enough for classifying meteorites, while the ICPs, especially ICP-MS, can characterize elemental abundance features in detail, as demonstrated by REE abundance patterns for the Allende meteorite. In this manner, INAA and ICPs can be used in a complementary fashion in cosmochemical studies.
ISSN:0236-5731
1588-2780
DOI:10.1007/s10967-018-6309-z