O-077 Measurement Of Respiratory Mechanics And Thoracoabdominal Asynchrony Indicies In Neonates By Respiratory Inductance Plethysmography During Non-invasive Ventilation

BackgroundMeasurement of respiratory mechanics during non-invasive ventilation (NIV) precludes use of the traditional airway flow sensor. Increasing use of NIV in premature infants necessitated novel instrumentation for measuring airflow without interfering with the nasal/oral interface. Respiratory...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Archives of disease in childhood 2014-10, Vol.99 (Suppl 2), p.A52-A53
Hauptverfasser: Gerdes, S, Sivieri, E, Abbasi, S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BackgroundMeasurement of respiratory mechanics during non-invasive ventilation (NIV) precludes use of the traditional airway flow sensor. Increasing use of NIV in premature infants necessitated novel instrumentation for measuring airflow without interfering with the nasal/oral interface. Respiratory inductance plethysmography (RIP), in addition to providing chest wall motion analysis, may be used for volume and airflow measurements when properly calibrated.ObjectiveTo develop an efficient RIP calibration technique to allow bedside measurement of respiratory mechanics and to validate it’s accuracy against traditional pneumotachometer (PNT) measurements while simultaneously computing thoracoabdominal asynchrony indices in premature infants.Design/methodsRIP ribcage and abdominal signals were recorded simultaneously with facemask PNT signals. RIP was calibrated by qualitative diagnostic calibration and multiple linear regression algorithms on artefact-free breaths with amplitudes within the mean±1SD. A sequence of optimising iterations for computer breath selection produced best-fit regression coefficients with the PNT flow. Transpulmonary pressure was measured by esophageal catheter. Lung compliance and airway resistance were computed by a least mean square technique. Ribcage to abdominal phase angle (phi), Laboured breathing index (LBI), and phase relation in total breath (PhRTB) were computed from analysis of RIP ribcage and abdomen excursions. Validation measurements were performed on 18 infants of 28–35wks GA.ResultsCorrelation coefficients for compliance and resistance by PNT vs. RIP were r2=0.9737 and 0.8980 respectively. LBI for these infants was 1.2 ± 0.6, phi was 54.4 ± 7.4 degrees and PhRTB was 33.4 ± 3.1%.ConclusionWhen properly calibrated, RIP derived respiratory mechanics measurements provide sufficient diagnostic accuracy in infants receiving NIV support.Abstract O-077 Figure 1
ISSN:0003-9888
1468-2044
DOI:10.1136/archdischild-2014-307384.144