Lesion dimensions during temperature-controlled radiofrequency catheter ablation of left ventricular porcine myocardium: Impact of ablation site, electrode size, and

BACKGROUND: It is important to increase lesion size to improve the success rate for radiofrequency ablation of ischemic ventricular tachycardia. This study of radiofrequency ablation, with adjustment of power to approach a preset target temperature, ie, temperature-controlled ablation, explores the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Circulation (New York, N.Y.) N.Y.), 1999-01, Vol.99 (2), p.319
Hauptverfasser: Helen Hogh Petersen, Chen, Xu, Pietersen, Adrian, Jesper Hastrup Svendsen, Haunso, Stig
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BACKGROUND: It is important to increase lesion size to improve the success rate for radiofrequency ablation of ischemic ventricular tachycardia. This study of radiofrequency ablation, with adjustment of power to approach a preset target temperature, ie, temperature-controlled ablation, explores the effect of catheter-tip length, ablation site, and convective cooling on lesion dimensions. METHODS AND RESULTS: In vitro strips of porcine left ventricular myocardium during different levels of convective cooling and in vivo pig hearts at 2 or 3 left ventricular sites were ablated with 2- to 12-mm-tip catheters. We found increased lesion volume for increased catheter-tip length