Energy dispersive X-ray fluorescence (EDXRF) equipment calibration for multielement analysis of soil and rock samples

Considering the importance to understand the behavior of the elements on different natural and/or anthropic processes, this study had as objective to verify the accuracy of a multielement analysis method for rocks characterization by using soil standards as calibration reference. An EDXRF equipment...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: de Moraes Alex Silva, Lohane, Tech, Melquíades, Fábio Luiz, Bastos, Rodrigo Oliveira
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Considering the importance to understand the behavior of the elements on different natural and/or anthropic processes, this study had as objective to verify the accuracy of a multielement analysis method for rocks characterization by using soil standards as calibration reference. An EDXRF equipment was used. The analyses were made on samples doped with known concentration of Mn, Zn, Rb, Sr and Zr, for the obtainment of the calibration curves, and on a certified rock sample to check the accuracy of the analytical curves. Then, a set of rock samples from Rio Bonito, located in Figueira city, Paraná State, Brazil, were analyzed. The concentration values obtained, in ppm, for Mn, Rb, Sr and Zr varied, respectively, from 175 to 1084, 7.4 to 268, 28 to 2247 and 15 to 761.
ISSN:0094-243X
1551-7616
DOI:10.1063/1.4901775