Note: Low temperature calibration of a commercial differential scanning calorimeter

A method for a direct calibration of temperature for a commercial differential scanning calorimeter is presented. The advantage of this calibration method is that it can be done for any number of selected temperatures and is not dependent on standards of known melting points in the temperature range...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2018-01, Vol.89 (1), p.016105-016105
1. Verfasser: Berberian, John G.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for a direct calibration of temperature for a commercial differential scanning calorimeter is presented. The advantage of this calibration method is that it can be done for any number of selected temperatures and is not dependent on standards of known melting points in the temperature range of interest. The calibration presented in this note is for the temperature range from 91 K to 273 K, the temperature range of interest, but the range can easily be extended.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.5005817