Development of an 8 channel sEMG wireless device based on ADS1299 with Virtual Instrumentation
In this paper, a different approach on the use of the ADS1299 (an analog front-end with features for electroencephalogram and electrocardiography signal acquisition) is considered, proposing the development of a surface electromyography (sEMG) device. The main features of the device include simultan...
Gespeichert in:
Veröffentlicht in: | arXiv.org 2018-08 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this paper, a different approach on the use of the ADS1299 (an analog front-end with features for electroencephalogram and electrocardiography signal acquisition) is considered, proposing the development of a surface electromyography (sEMG) device. The main features of the device include simultaneous recordings of eight muscular channels, wireless transmission and virtual instrumentation with the use of LabVIEWTM software. The proposed sEMG device contains a specifically designed protocol to accommodate data transmission by reducing the data size while still delivering adequate resolution (34.33 uV), amplitude range (17.57 mV) and sampling rate (1000 Hz) for sEMG signals. For the validation methods, a generated sine wave and a known sEMG data were evaluated. Moreover, the muscular recordings for all the eight channels of a human arm were successful and the results expose the isolated contractions of the triceps and the biceps with their amplitude range and frequency spectrum. |
---|---|