A liquid-helium cooled large-area silicon PIN photodiode x-ray detector
An x-ray detector using a liquid-helium cooled large-area silicon PIN photodiode has been developed along with a tailor-made charge sensitive preamplifier whose first-stage JFET has been cooled. The operating temperature of the JFET has been varied separately and optimized. The x- and \(\gamma\)-ray...
Gespeichert in:
Veröffentlicht in: | arXiv.org 1995-04 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An x-ray detector using a liquid-helium cooled large-area silicon PIN photodiode has been developed along with a tailor-made charge sensitive preamplifier whose first-stage JFET has been cooled. The operating temperature of the JFET has been varied separately and optimized. The x- and \(\gamma\)-ray energy spectra for an \nuc{241}{Am} source have been measured with the photodiode operated at 13 K. An energy resolution of 1.60 keV (FWHM) has been obtained for 60-keV \(\gamma\) rays and 1.30 keV (FWHM) for the pulser. The energy threshold could be set as low as 3 keV. It has been shown that a silicon PIN photodiode serves as a low-cost excellent x-ray detector which covers large area at 13 K. |
---|---|
ISSN: | 2331-8422 |
DOI: | 10.48550/arxiv.9504010 |