1550nm fiber optic TOAD (time of arrival diagnostic) for measuring sub-nanosecond resolution of detonation break out
We report on the design and testing of an eight-channel 1550 nm TOAD (Time Of Arrival Diagnostic) system that we have shown to resolve the arrival of a detonation wave front on explosive dynamic experiments with a timing resolution of ±240ps. The TOAD probe consists of a 1550 nm single mode optical...
Gespeichert in:
Hauptverfasser: | , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report on the design and testing of an eight-channel 1550 nm TOAD (Time Of Arrival Diagnostic) system that we have shown to resolve the arrival of a detonation wave front on explosive dynamic experiments with a timing resolution of ±240ps. The TOAD probe consists of a 1550 nm single mode optical fiber, which is mounted, in a flat polished ferrule. The exposed optical fiber is coated with 1000 angstroms of aluminum. We have demonstrated that the TOAD system can measure detonation arrival times on metal or bare high explosives. By omitting the 1000 angstroms of aluminum on the end of the single mode optical fiber, the 1550 nm TOAD system can be converted into a very simple PDV (Photon Doppler Velocimetry) diagnostic. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.5045024 |