Comparison of different methods for evaluating the transmission function of a two-stage cylindrical mirror analizer in XPS applications
Three different evaluations of the energy dependence of the transmission-detection function of a two-stage cylindrical electron analyzer have been obtained by resorting to three established procedures. Their relative merits have been tested as follows. First they have been used to correct raw XPS sp...
Gespeichert in:
Veröffentlicht in: | arXiv.org 2010-01 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Three different evaluations of the energy dependence of the transmission-detection function of a two-stage cylindrical electron analyzer have been obtained by resorting to three established procedures. Their relative merits have been tested as follows. First they have been used to correct raw XPS spectra of clean Cu, Ag and Au surfaces. Next, the secondary electron background has been subtracted using Tougaard's method. Finally, the primary electron spectra so obtained have been reanalyzed by peak area measurement in the frame of the modern formalism for quantitative XPS analysis. Ideally a constant residual value should thus be obtained. The variability of these residuals with peak energy allows an objective rating of the initial transmission-detection functions. |
---|---|