A Batteryless Energy Harvesting Storage System for Implantable Medical Devices Demonstrated In Situ

We report a wireless energy harvesting and telemetry storage system in 180 nm CMOS technology, demonstrated in situ in rat carcass. The implantable device has dimensions 13 mm × 15 mm and stores 87.5 mJ, providing a self-powering time of 8.5 s transmitting through tissue. We utilize an all-solid-sta...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Circuits, systems, and signal processing systems, and signal processing, 2019-03, Vol.38 (3), p.1360-1373
Hauptverfasser: Gall, Oren Z., Meng, Chuizhou, Bhamra, Hansraj, Mei, Henry, John, Simon W. M., Irazoqui, Pedro P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report a wireless energy harvesting and telemetry storage system in 180 nm CMOS technology, demonstrated in situ in rat carcass. The implantable device has dimensions 13 mm × 15 mm and stores 87.5 mJ, providing a self-powering time of 8.5 s transmitting through tissue. We utilize an all-solid-state flexible supercapacitor of breakdown voltage 0.8 V and capacitance 400 mF to harvest incoming wireless power, followed by a boost converter CMOS that drives an active wireless transmitter at 1.5 V at 2.4 GHz in the industrial, scientific, and medical (ISM) band. The DC/DC converter component and switching frequency selection were guided by genetic algorithm analysis and use digital feedback to control the pulse width modulation (PWM), which slowly modifies the duty cycle to control output voltage fluctuations. This implantable medical device system presents the roadmap for batteryless energy harvesting in vivo and in clinical environments, exhibiting the highest operating storage density of 450 μJ/mm 2 reported to date.
ISSN:0278-081X
1531-5878
DOI:10.1007/s00034-018-0915-4