A Fluorimeter on the Basis of a Femtosecond Cr+4: Forsterite Laser
An apparatus for measuring luminescence lifetimes using the femtosecond up-conversion method is presented. A laser system based on a chromium–forsterite crystal with a lasing wavelength of 1250 nm is used to generate ultrashort pulses. Luminescence is excited in a sample by optical harmonics that ar...
Gespeichert in:
Veröffentlicht in: | Instruments and experimental techniques (New York) 2018-07, Vol.61 (4), p.556-561 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An apparatus for measuring luminescence lifetimes using the femtosecond up-conversion method is presented. A laser system based on a chromium–forsterite crystal with a lasing wavelength of 1250 nm is used to generate ultrashort pulses. Luminescence is excited in a sample by optical harmonics that are generated in nonlinear optical crystals. The width of the instrument function of the system is 200 fs. Methods for optimizing the signal-to-noise ratio in measurements are discussed. The results of measurements of the fluorescence- decay dynamics and the anisotropy of aqueous solutions of an inclusion complex of the 4-DASPI styrene dye in cucurbit[6]uril on short (5 ps) and long (500 ps) time scales are presented. |
---|---|
ISSN: | 0020-4412 1608-3180 |
DOI: | 10.1134/S0020441218040103 |