Evaluation of dose rate and photon energy dependence of PASSAG polymer gel dosimeter

In the current study, dose rate and photon energy dependence of PASSAG gel dosimeter is evaluated by using MRI technique. The gel samples were readies under normal atmospheric conditions and then were irradiated after 24 h with 100, 200, 300, and 400 cGy/min dose rates and photon energies of 6 and 1...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radioanalytical and nuclear chemistry 2018-08, Vol.317 (2), p.1041-1050
Hauptverfasser: Farhood, Bagher, Geraily, Ghazale, Abtahi, Seyed Mohammad Mahdi, Ghorbani, Mahdi, Mehdikhani, Mohamad
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In the current study, dose rate and photon energy dependence of PASSAG gel dosimeter is evaluated by using MRI technique. The gel samples were readies under normal atmospheric conditions and then were irradiated after 24 h with 100, 200, 300, and 400 cGy/min dose rates and photon energies of 6 and 18 MV from a linear accelerator. The results demonstrate that there is no considerable effect of dose rate and photon energy dependence on the R 2 -dose response and sensitivity of PASSAG gel dosimeter in 0–10 Gy dose range, as there are 0.01–2.80 and 1.46–2.52% difference ranges between the R 2 values of the gel samples at four different dose rates and two various photon energies, respectively.