Accuracy of Determining Coordinates by Means of Charge Division in a Position-Sensitive Thermal Neutron Detector Based on Boron-10

A charge division algorithm is proposed for determining coordinates in a positional sensitive neutron detector based on a layer of 10 B and an ion chamber. Experimental impedances of the input resistances of amplifiers and their gain factors are considered in order to improve spatial resolution. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Bulletin of the Russian Academy of Sciences. Physics 2018-06, Vol.82 (6), p.748-751
Hauptverfasser: Karaevsky, S. Kh, Potashev, S. I., Drachev, A. I., Burmistrov, Yu. M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A charge division algorithm is proposed for determining coordinates in a positional sensitive neutron detector based on a layer of 10 B and an ion chamber. Experimental impedances of the input resistances of amplifiers and their gain factors are considered in order to improve spatial resolution. The best spatial resolution of ~2 mm with respect to the horizontal X-coordinate and ~4 mm with respect to the vertical Y-coordinate is achieved at a power supply voltage of 700 V. The efficiency of thermal neutron registration is ~3%. A helium-3 counter of known efficiency is used to monitor thermal neutrons.
ISSN:1062-8738
1934-9432
DOI:10.3103/S1062873818060151