99mTc-肝受容体シンチグラフィにおけるプレイナ画像とSPECT 画像の肝機能指標の精度の検証

Purpose: The purpose of this study was to evaluate the accuracy for quantitative predictor of hepatic functional reserve using planar and single photon emission computed tomography (SPECT) images. Methods: We acquired planar and SPECT images of the myocardial torso phantom and calculated LHL (uptake...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nippon Hōshasen Gijutsu Gakkai zasshi 2017, Vol.73(10), pp.1055-1060
Hauptverfasser: 長谷川, 大輔, 大西, 英雄
Format: Artikel
Sprache:jpn
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: The purpose of this study was to evaluate the accuracy for quantitative predictor of hepatic functional reserve using planar and single photon emission computed tomography (SPECT) images. Methods: We acquired planar and SPECT images of the myocardial torso phantom and calculated LHL (uptake ratio of the liver to the liver plus heart). Planar images were used with and without scatter correction. Myocardial torso phantom was simulated radioactivity distribution of normal liver function and poor liver function. This study compared the LHLplanar (calculated by planar images) and the LHLSPECT (calculated by SPECT images) with the true value (calculated by radioactivity concentration in the phantom) and evaluated the error rate. Results: Error rate of LHLSPECT was improved compared with the LHLplanar. The error rate of LHLplanar (with scatter correction) was increased in the poor liver function than in the normal liver function. Conclusions: The SPECT images have higher advantage to predict the liver function than the planar images.
ISSN:0369-4305
1881-4883
DOI:10.6009/jjrt.2017_JSRT_73.10.1055