APPARATUS: A randomised, crossover study of the Dorges, McCoy and Macintosh laryngoscope blades in a simulated difficult intubation scenario

The use of alternative adult laryngoscope blades in an unexpected difficult intubation was evaluated with a human patient simulator manikin. Twenty anaesthetists of varying experience attempted to intubate the trachea of a high fidelity simulator in both its normal and difficult intubation settings...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 2006-05, Vol.61 (5), p.482
Hauptverfasser: Sethuraman, D, Darshane, S, Guha, A, Charters, P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The use of alternative adult laryngoscope blades in an unexpected difficult intubation was evaluated with a human patient simulator manikin. Twenty anaesthetists of varying experience attempted to intubate the trachea of a high fidelity simulator in both its normal and difficult intubation settings using Macintosh, Dorges and McCoy laryngoscopes in a randomised order. The time taken to intubate, Cormack and Lehane scores, percentage of glottic opening visible, failure rate, number of attempts and subjective ease of use were recorded. The Dorges and McCoy blades did not perform any better than the standard Macintosh blade in either the easy or difficult tracheal intubation settings. Guidelines recommending the use of an alternative blade in an unexpected difficult intubation scenario have limited supporting evidence. This study does not support this recommendation.[PUBLICATION ABSTRACT]
ISSN:0003-2409
1365-2044
DOI:10.1111/j.1365-2044.2006.04613.x