Radiocarbon Analysis on the New AARAMS 1MV Tandetron
We report here on the radiocarbon performance on the AARAMS HVE 1MV Tandetron. 14C analysis is carried out in charge state 2+. We have avoided Li interference by appropriate settings of the high-energy electrostatic analyzer and the 30° second high-energy magnet. The 14C machine background is determ...
Gespeichert in:
Veröffentlicht in: | Radiocarbon 2017-06, Vol.59 (3), p.905-913 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report here on the radiocarbon performance on the AARAMS HVE 1MV Tandetron. 14C analysis is carried out in charge state 2+. We have avoided Li interference by appropriate settings of the high-energy electrostatic analyzer and the 30° second high-energy magnet. The 14C machine background is determined using unprocessed graphite, which yielded 58,650±2032 14C yr determined as the average and standard deviation of four measurements. International standards, which are used to monitor the long-term performance of the 14C measurements, agree with the reported consensus values. |
---|---|
ISSN: | 0033-8222 1945-5755 |
DOI: | 10.1017/RDC.2016.85 |