Time constant optimization of solar irradiance absolute radiometer

We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer(SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06%...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optoelectronics letters 2017-05, Vol.13 (3), p.179-183
1. Verfasser: 唐潇 方伟 王玉鹏 杨东军 衣小龙
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer(SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06% in the midday and 0.275% in the morning and afternoon. The uncertainty level which can be considered negligible for SIAR is also investigated, and it is suggested that the uncertainty level has to be less than 0.02%. Then, combining the requirement of international comparison with these two conclusions, we conclude that the suitable time constant for SIAR is 20 s.
ISSN:1673-1905
1993-5013
DOI:10.1007/s11801-017-7032-7