Correction of output from an internal load cell in a high-pressure triaxial deformation apparatus without a split-piston
Internal load cells can be used to measure the axial load applied to a sample during a triaxial experiment precisely, which is key to obtaining reliable mechanical data. However, the output from an internal load cell contains an offset, which is a function of the confining pressure, pore pressure, a...
Gespeichert in:
Veröffentlicht in: | Chishitsugaku zasshi 2016/12/15, Vol.122(12), pp.653-658 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Internal load cells can be used to measure the axial load applied to a sample during a triaxial experiment precisely, which is key to obtaining reliable mechanical data. However, the output from an internal load cell contains an offset, which is a function of the confining pressure, pore pressure, and temperature. For an apparatus without a split-piston, the sample column has a non-zero axial load supported by O-rings even before a hit-point. Thus, a simple zero setting before the hit-point would cause an error in the measured axial load. In this technical note, we present a method that can be used to estimate the offset and correct the output by using outputs from both internal and external load cells. This method does not require additional experimentation. |
---|---|
ISSN: | 0016-7630 1349-9963 |
DOI: | 10.5575/geosoc.2016.0047 |