Correction of output from an internal load cell in a high-pressure triaxial deformation apparatus without a split-piston

Internal load cells can be used to measure the axial load applied to a sample during a triaxial experiment precisely, which is key to obtaining reliable mechanical data. However, the output from an internal load cell contains an offset, which is a function of the confining pressure, pore pressure, a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chishitsugaku zasshi 2016/12/15, Vol.122(12), pp.653-658
Hauptverfasser: Noda, Hiroyuki, Takahashi, Miki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Internal load cells can be used to measure the axial load applied to a sample during a triaxial experiment precisely, which is key to obtaining reliable mechanical data. However, the output from an internal load cell contains an offset, which is a function of the confining pressure, pore pressure, and temperature. For an apparatus without a split-piston, the sample column has a non-zero axial load supported by O-rings even before a hit-point. Thus, a simple zero setting before the hit-point would cause an error in the measured axial load. In this technical note, we present a method that can be used to estimate the offset and correct the output by using outputs from both internal and external load cells. This method does not require additional experimentation.
ISSN:0016-7630
1349-9963
DOI:10.5575/geosoc.2016.0047