Influence of temperature and turbidity on water COD detection by UV absorption spectroscopy

Ultraviolet(UV) absorption spectroscopy is used to detect the concentration of water chemical oxygen demand(COD). The UV absorption spectra of COD solutions are analyzed qualitatively and quantitatively. The partial least square(PLS) algorithm is used to model COD solution and the modeling results a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Optoelectronics letters 2016-11, Vol.12 (6), p.461-464
1. Verfasser: 周昆鹏 毕卫红 张启航 付兴虎 吴国庆
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Ultraviolet(UV) absorption spectroscopy is used to detect the concentration of water chemical oxygen demand(COD). The UV absorption spectra of COD solutions are analyzed qualitatively and quantitatively. The partial least square(PLS) algorithm is used to model COD solution and the modeling results are compared. The influence of environmental temperature and turbidity is analyzed. These results show that the influence of temperature on the predicted value can be ignored. However, the change of turbidity can affect the detection results of UV spectra, and the COD detection error can be effectively compensated by establishing the single-element regression model.
ISSN:1673-1905
1993-5013
DOI:10.1007/s11801-016-6178-z