DUAL-AXIS ROTATIONAL CORONARY ANGIOGRAPHY CAN REDUCE PEAK SKIN DOSE AND SCATTERED DOSE: A PHANTOM STUDY
Objectives to evaluate the peak skin dose received by the patient and scattered dose to the operator during the dual-axis rotational coronary angiography and compared to standard coronary angiography. Methods An anthropomorphic phantom was used to simulate coronary angiography and generate radiation...
Gespeichert in:
Veröffentlicht in: | Heart (British Cardiac Society) 2012-10, Vol.98 (Suppl 2), p.E291-E291 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives to evaluate the peak skin dose received by the patient and scattered dose to the operator during the dual-axis rotational coronary angiography and compared to standard coronary angiography. Methods An anthropomorphic phantom was used to simulate coronary angiography and generate radiation dose. Procedures included the simulation of standard coronary angiography taking into account the complexity of the individual lesion and dual-axis rotational coronary angiography. The peak skin dose was measured with thermoluminescent dosimeter arrays and scattered dose with a dosimeter at predefined height (approximately at the level of left arm) at the operator's location. Results Measurements showed, compared to standard coronary angiography, there is a reduction in peak skin dose (the overall reduction: 68%) and scattered dose (the overall reduction: 58%) with the use of dual-axis rotational coronary angiography. Moreover, these reductions become greater in the set of complicated lesion examinations (a reduction of 82% for peak skin dose and 70% for scattered dose). Conclusions In comparison to standard coronary angiography, dual-axis rotational coronary angiography can reduce peak skin dose and scattered dose, especially for the complicated lesion examinations. The major reasons seem to be the overall decreased amount of applied energy during the dual-axis rotational coronary angiography, and that it can spread the administered dose over a larger area and minimise the peak skin dose. |
---|---|
ISSN: | 1355-6037 1468-201X |
DOI: | 10.1136/heartjnl-2012-302920ad.5 |