116 CRT optimisation: improving echocardiographic techniques by accommodating biological variability within different echocardiographic parameters

BackgroundIn optimisation of CRT (and even selection for implantation) we may have underestimated the impact of beat-to-beat variation on echocardiographic measurements. This can be quantified most clearly in the optimisation process, in which genuine small changes in cardiac function (signal) must...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart (British Cardiac Society) 2011-06, Vol.97 (Suppl 1), p.A66-A67
Hauptverfasser: Pabari, P A, Kyriacou, A, Moraldo, M, Unsworth, B, Sutaria, N, Mayet, J, Hughes, A D, Francis, D P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BackgroundIn optimisation of CRT (and even selection for implantation) we may have underestimated the impact of beat-to-beat variation on echocardiographic measurements. This can be quantified most clearly in the optimisation process, in which genuine small changes in cardiac function (signal) must be detected among potentially large beat-to-beat variation (noise).Methods and ResultsIn this large study of biological variability, we performed over 2000 echocardigraphic measurements in 12 patients. We performed separate, replicate measurements at a series of interventricular delays of each potential optimisation modality at rest. This included (i) 3D systolic dyssynchrony index, (ii) aortic pre-ejection time, (iii) interventricular mechanical delay, (iv) LVOT VTI and (v) QRS width. The equivalent of 31 optimisations per patient were performed. For single measurements at each setting, agreement between successive optimisations was low, at 39% for SDI, 41% for aortic pre-ejection time, 32% for IVMD, 54% for LVOT VTI and 58% for QRS width. Agreement between one method and another, using single replicates, was similarly low, with the average agreement between optima by two methods being only 18% similar to pure guesswork. The intraclass correlation coefficient was low for all methods at 0.11, 0.51, 0.30, 0.50 and 0.55 respectively. The intraclass correlation coefficients improved to 0.19, 0.63, 0.42, 0.54 and 0.66 (p=0.001) when averages of paired measurements were used. To optimise within 20 ms or 10 ms of the true optimum, requires a greater number of measurements, as seen in Abstract 116 figure 1, dependant on the intraclass correlation coefficient. The scatter of optima obtained reduced (improved) significantly when using averaged pairs of measurements compared to single measurements from 23 ms to 18 ms (3D SDI), 14 ms to 10 ms (aortic pre-ejection time), 28 ms to 22 ms (IVMD), 21 ms to 16 ms (LVOT VTI) and 14 ms to 10 ms QRS duration (p=0.0002).Abstract 116 Figure 1ConclusionsBecause of beat-to-beat variability, VV delay optimisation by any of the echocardiographic techniques is not realistic unless multiple replicates are performed and averaged. Smoothing biological variation by averaging multiple measurements allows the full potential of echocardiographic optimisation to be achieved and improves the consistency of optimisation. Trying to save time by performing inadequate numbers of replicates is a false economy and leads to optimisation being a form of r
ISSN:1355-6037
1468-201X
DOI:10.1136/heartjnl-2011-300198.116