TCT-526 An in-vivo multi-modality imaging study of the Absorb everolimus-eluting bioresorbable scaffold in complex coronary bifurcations

Bifurcation stenting procedures of the aorta-iliac bifurcation (70° angle) were performed with 3.0x28 mm BVS using the following techniques: main-vessel (MV) stenting with ballooning of side branch (SB) through the BVS struts (Provisional stenting, n=5), T-and protrusion (TAP, n=5), modified T (n=5)...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American College of Cardiology 2015-10, Vol.66 (15), p.B215-B215
Hauptverfasser: Bennett, Johan, Driessche, Nina Vanden, Vanhaverbeke, Maarten, Desmet, Walter, Sinnaeve, P, Adriaenssens, Tom, Dubois, Christophe
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Bifurcation stenting procedures of the aorta-iliac bifurcation (70° angle) were performed with 3.0x28 mm BVS using the following techniques: main-vessel (MV) stenting with ballooning of side branch (SB) through the BVS struts (Provisional stenting, n=5), T-and protrusion (TAP, n=5), modified T (n=5) and culotte (n=5) stenting. [...]culotte stenting frequently caused significant circumferential malapposition, scaffold distortion and strut fractures, the clinical impact of which is unknown.
ISSN:0735-1097
1558-3597
DOI:10.1016/j.jacc.2015.08.543