Comparison between the M and XL probes for liver fibrosis assessment by transient elastography
Liver stiffness measurement (LSM) using Transient Elastography (TE) for liver fibrosis assessment is difficult to be performed in obese and overweight patients by standard M probe, thus the XL probe was developed. The aim of our paper was to assess the usefulness of the XL probe in daily clinical pr...
Gespeichert in:
Veröffentlicht in: | Medical ultrasonography 2014-06, Vol.16 (2), p.119-122 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Liver stiffness measurement (LSM) using Transient Elastography (TE) for liver fibrosis assessment is difficult to be performed in obese and overweight patients by standard M probe, thus the XL probe was developed. The aim of our paper was to assess the usefulness of the XL probe in daily clinical practice.
Our study included 216 patients (mean BMI 30.1+/-4.1 kg/m2) with chronic hepatopathies, in which paired measurements were made using the M (3.5MHz) and XL (2.5 MHz) probes in the same session. In each patient 10 valid LSM were acquired with each probe, a median was calculated, expressed in kiloPascals (kPa). Unreliable TE measurements were considered: fewer than 10 valid shots; with a success rate (SR) |
---|---|