An investigation of rotational issues for rectal carcinoma treated with radiotherapy
Abstract Purpose To quantify the amount of inter-fractional pitch for rectal carcinoma patients, to investigate the dosimetric impact of pitch on the target volume and critical structures and to determine a tolerance where no pitch correction is required. Materials and methods Daily pre-treatment im...
Gespeichert in:
Veröffentlicht in: | Journal of radiotherapy in practice 2014-06, Vol.13 (2), p.195-202 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Purpose To quantify the amount of inter-fractional pitch for rectal carcinoma patients, to investigate the dosimetric impact of pitch on the target volume and critical structures and to determine a tolerance where no pitch correction is required. Materials and methods Daily pre-treatment images of rectal carcinoma patients were analysed to determine the residual pitch compared with the computed tomography (CT) planning scan. The dosimetric impact of pelvic rotation was modelled. The dose coverage of the clinical target volume (CTV) and small bowel were evaluated using dose-volume histograms. Results Pre-treatment images had a mean of 0·27° and standard deviation was 2·23°. The volume of CTV receiving 95% of the prescription dose altered by 0·1% when up to ±10° of pitch was simulated. Conclusions No clinically significant change in CTV coverage was found (when ±10° of pitch was simulated). A tolerance of ±10° of pitch has been implemented for rectal carcinoma patients treated with three-dimensional conformal radiotherapy in our institution, when daily pre-treatment imaging with a zero action threshold for translational shifts is used. |
---|---|
ISSN: | 1460-3969 1467-1131 |
DOI: | 10.1017/S146039691300023X |