Electrochemical Detector for High-Performance Liquid Chromatography. VI. Application of a Twin Electrochemical Detector
The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simulta...
Gespeichert in:
Veröffentlicht in: | Chemical & pharmaceutical bulletin 1983/10/25, Vol.31(10), pp.3571-3578 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simultaneous determination of reduced and oxidized species is possible, 2) the stability of the electrode is improved by applying a larger potential to the other electrode, 3) the specificity of detection is improved by the re-electrolysis at the second electrode of the reaction product produced at the first electrode, and 4) specific detection of the post-discharging materials and sensitive detection are possible by using appropriate operating circuits. |
---|---|