Electrochemical Detector for High-Performance Liquid Chromatography. VI. Application of a Twin Electrochemical Detector

The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simulta...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chemical & pharmaceutical bulletin 1983/10/25, Vol.31(10), pp.3571-3578
Hauptverfasser: HIROSHIMA, OSAMU, IKENOYA, SATORU, NAITOH, TOSHIHIKO, KUSUBE, KENICHI, OHMAE, MASAHIKO, KAWABE, KIYOSHI, ISHIKAWA, SEIJI, HOSHIDA, HARUHIKO, KURAHASHI, TATSUO
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simultaneous determination of reduced and oxidized species is possible, 2) the stability of the electrode is improved by applying a larger potential to the other electrode, 3) the specificity of detection is improved by the re-electrolysis at the second electrode of the reaction product produced at the first electrode, and 4) specific detection of the post-discharging materials and sensitive detection are possible by using appropriate operating circuits.