Evaluation of colorimetric method for zinc, using clinical chemistry analyzer
We developed a colorimetric reagent kit for determination of zinc in serum, “ACCURAS AUTO Zn”. The reagent is applicable to all auto-analyzers widely used in hospital laboratories, without any serum pretreatment. Within-run and between-run precisions(C. V. )were 0.7-1.0% and 1.4-1.9%, respectively....
Gespeichert in:
Veröffentlicht in: | Biomedical Research on Trace Elements 2007/12/31, Vol.18(4), pp.380-385 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng ; jpn |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We developed a colorimetric reagent kit for determination of zinc in serum, “ACCURAS AUTO Zn”. The reagent is applicable to all auto-analyzers widely used in hospital laboratories, without any serum pretreatment. Within-run and between-run precisions(C. V. )were 0.7-1.0% and 1.4-1.9%, respectively. The calibration curve was linear up to 500 µg/dl, and the detection limit was 5 µg/dl. Interference from iron and copper was eliminated by a masking agent. A good correlation(r = 0.996)was obtained between the results of the present method and those of atomic absorption spectrophotometry. This kit enables accurate and immediate assay of zinc with a clinical chemistry auto-analyzer, and saves time and efforts in the field of a nutrition support team(NST)work in hospitals. |
---|---|
ISSN: | 0916-717X 1880-1404 |
DOI: | 10.11299/brte.18.380 |