On the Calibration of Optical Full-Field Strain Measurement Systems
There are no standard reference materials suitable for the calibration of full-field optical strain measurement systems. This is hindering the uptake of the technology by industrial end-users since optical metrology instrumentation and procedures cannot be easily integrated into quality assurance sy...
Gespeichert in:
Veröffentlicht in: | Applied Mechanics and Materials 2006-08, Vol.3-4, p.397-402 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | There are no standard reference materials suitable for the calibration of full-field optical strain measurement systems. This is hindering the uptake of the technology by industrial end-users since optical metrology instrumentation and procedures cannot be easily integrated into quality assurance systems. The EU-funded SPOTS project is developing a physical reference material (PRM) and measurement protocol that should provide the basis of a calibration standard for establishing the traceability of strain values obtained with optical devices. This paper describes a PRM based on a parametric design of monolithic four-point bend test that can reliably generate a known strain field over a range of specimen sizes. Measurements acquired from strain gauges and LVDTs compared well with data obtained from ESPI, digital image correlation, photoelasticity and thermoelasticity studies, demonstrating excellent repeatability and inter-laboratory reproducibility. |
---|---|
ISSN: | 1660-9336 1662-7482 1662-7482 |
DOI: | 10.4028/www.scientific.net/AMM.3-4.397 |