Hidden dangers in a simulation centre
Abstract In the article, we emphasize the need for all equipment in a simulation centre to be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve wen...
Gespeichert in:
Veröffentlicht in: | Zdravniški vestnik (Ljubljana, Slovenia : 1992) Slovenia : 1992), 2011-06, Vol.80 (6) |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | slv |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract In the article, we emphasize the need for all equipment in a simulation centre to be maintained in excellent working condition, and present a case of nitrous oxide (N2O) leakage through a damaged pressure regulator. For a considerable time faulty operation of the pressure-reducing valve went unnoticed, resulting in slow, imperceptible leakage of gas from the cylinder. A careful soap test revealed that gas was escaping through a small nozzle on the back of the pressure regulator. The leakage was inaudible but could be felt by placing a finger on the valve nozzle. |
---|---|
ISSN: | 1318-0347 1581-0024 |