Apparatus for Measuring Thermal Conductivity of Ceramic and Metallic Materials to 1200°C

A steady‐state apparatus for measuring thermal conductivity by a direct method, viz., water calorimetry, is described. The accuracy of measurement is about 1 to 3% in the temperature range 30° to 30O°C. and about 3 to 10% in the range 300° to 1200°C. The samples are rods 9 in. long and in. in diamet...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American Ceramic Society 1960-02, Vol.43 (2), p.81-85
1. Verfasser: SUTTON, W. H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A steady‐state apparatus for measuring thermal conductivity by a direct method, viz., water calorimetry, is described. The accuracy of measurement is about 1 to 3% in the temperature range 30° to 30O°C. and about 3 to 10% in the range 300° to 1200°C. The samples are rods 9 in. long and in. in diameter, although smaller sizes can be accommodated. One feature of the design is that the sample thermocouples can be plugged into the unit. Curves for samples of alumina, stainless steel, and graphite are included.