Technical evaluation of dialysate flow in a hollow-fiber dialyzer

In a hollow-fiber dialyzer, uremic toxins are removed by diffusion and convection, which are influenced by the dialysate flow patterns in the dialyzer. Recently available high-performance dialyzers have complicated dialysate flow patterns, because both positive filtration and negative filtration occ...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of artificial organs 2002-12, Vol.5 (4), p.251-256
Hauptverfasser: Ishiwata, K., Yamamoto, K., Kohori, F., Sakai, K., Fukuda, M., Hiyoshi, T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In a hollow-fiber dialyzer, uremic toxins are removed by diffusion and convection, which are influenced by the dialysate flow patterns in the dialyzer. Recently available high-performance dialyzers have complicated dialysate flow patterns, because both positive filtration and negative filtration occur. The objective of the present study was to evaluate dialysate flow in high-performance dialyzers experimentally. Glass-coated 0.1mmΦ platinum electrodes were used for the electrode counter and the working electrode. A counter electrode was placed at the inlet of the dialyzer, and working electrodes were placed at 20 different positions. A voltage of 0.5V was applied between the counter and the working electrodes with a potentiostat, and after the dialyzer was filled with water purified by reverse osmosis, 0.9% NaCl solution was caused to flow. The time at which the 0.9% NaCl solution reached each working electrode from the counter electrode was then measured at a dialysate-side flow rate of 300ml/min and blood-side flow rates of 0 and 200ml/min. It was found that in dialyzers with high permeability to pure water, dialysate flow was affected by both positive and negative filtration. A comparison was then made between the experimental results and the results of simulation by the finite element method; at positions at which positive and negative filtration occurred, good agreement was obtained. This method makes possible the experimental evaluation of dialysate flow in a high-performance dialyzer in which positive and negative filtration occur.[PUBLICATION ABSTRACT]
ISSN:1434-7229
1619-0904
DOI:10.1007/s100470200047