THE AMBIENT TEMPERATURE AND QUALITY OF THE HARDNESS MEASUREMENT

Three appraisers carried out the calibration of hardness tester by Vickers and Brinell methods, using on standard (CRM) with defined both values of standard hardness and their uncertainty. The ambient temperature of laboratory varied in range 14.9 to 27.6°C at ten levels. The influence of the temper...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of Faculty Engineering Hunedoara 2012-07, Vol.10 (3), p.379
Hauptverfasser: Petrík, Jozef, Miklos, Vojtech, Horváth, Matús
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Three appraisers carried out the calibration of hardness tester by Vickers and Brinell methods, using on standard (CRM) with defined both values of standard hardness and their uncertainty. The ambient temperature of laboratory varied in range 14.9 to 27.6°C at ten levels. The influence of the temperature on the results of calibration was evaluated by uncertainty, systematic and random errors whereby last two were determined by Youden plot. [PUBLICATION ABSTRACT]
ISSN:1584-2665
2601-2332