Reduction of contamination risks during clinical studies with Technegas

During patient studies with the Technegas equipment in our department, we regularly detected small to considerable contaminations of the operator and in the area surrounding the apparatus. These contaminations were found to be of different origin: residual activity in the tubing from the apparatus t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of nuclear medicine 1999-10, Vol.26 (10), p.1349-1352
Hauptverfasser: VANBILLOEN, H. P, BAUWENS, J, MORTELMANS, L, VERBRUGGEN, A. M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:During patient studies with the Technegas equipment in our department, we regularly detected small to considerable contaminations of the operator and in the area surrounding the apparatus. These contaminations were found to be of different origin: residual activity in the tubing from the apparatus to the patient which diffuses after deconnection, residual activity and small particles of the destroyed carbon crucible in the apparatus which are dispersed during opening of the door of the gas preparation chamber, leakage at the patient site during studies in uncooperative patients and a dysfunctioning valve inside the apparatus. To reduce the contamination risks, we made some adaptations to the apparatus. In the first place, the dysfunctioning valve was replaced. In addition, a powerful air exhaust pump with an efficient filter was installed. It was connected with (1) a newly installed transparent box in front of the door of the gas preparation chamber, (2) a dome on a flexible arm which can be positioned above the patient's face during the examination and (3) a nipple on which the mouthpiece can be placed after the study. After these adaptations, a study showed the absence of measurable contamination on the clothing of the personnel handling the apparatus. Occasionally, considerable contamination was still measured on the gloves worn during filling of the carbon boat with generator eluate, but only small contaminations (up to 9.25 kBq) were measured on the mouthmask worn by the operator during administration of the Technegas. This results in a maximum effective dose equivalent from activity deposited in the lungs of 0.008 microSv per study. The total body dose of the operator from external radiation for one Technegas examination was determined to be 2 microSv. The highest dose rate was measured during filling of the crucible (0.2 mSv/h).
ISSN:0340-6997
1619-7070
1619-7089
DOI:10.1007/s002590050594