Evaluation in mice of some iodinated melanoma imaging agents using cryosectioning and multi-wire proportional counting

AMBIS 4000, a multi-wire proportional counter, was calibrated for iodine-125 measurements. The detector displayed a linear response over a wide dynamic range. Using whole-body mice cryosections, a linear relationship could be established between count rate per area (cpm/mm2) measured with the AMBIS...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of nuclear medicine 1999-05, Vol.26 (5), p.494-498
Hauptverfasser: LABARRE, P, PAPON, J, MOREAU, M.-F, MOINS, N, VEYRE, A, MADELMONT, J.-C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:AMBIS 4000, a multi-wire proportional counter, was calibrated for iodine-125 measurements. The detector displayed a linear response over a wide dynamic range. Using whole-body mice cryosections, a linear relationship could be established between count rate per area (cpm/mm2) measured with the AMBIS 4000 detector and the count rate per gram (dpm/g) determined with an NaI(Tl) scintillation detector. A calibration curve could, thus be constructed. This new method allowed direct visual and quantitative evaluation of the biodistribution of a short series of 125I-labelled benzamides in melanoma-bearing mice. All the compounds studied showed good tumoral targeting ability. For one of them, ortho-N-(2-diethylaminoethyl)-4-iodobenzamide, liver and lung uptake decreased rapidly after dosing, making it a suitable tracer for scintigraphic detection of malignant melanoma.
ISSN:0340-6997
1619-7070
1619-7089
DOI:10.1007/s002590050416