Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy

This study evaluated the feasibility of utilizing a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance (QA) in intensity-modulated radiotherapy (IMRT). Contoured left and right head phantoms were converted from DICOM to STL format. Fused deposition modeli...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:PloS one 2017-07, Vol.12 (7), p.e0181560-e0181560
Hauptverfasser: Yea, Ji Woon, Park, Jae Won, Kim, Sung Kyu, Kim, Dong Youn, Kim, Jae Gu, Seo, Chan Young, Jeong, Won Hyo, Jeong, Man Youl, Oh, Se An
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study evaluated the feasibility of utilizing a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance (QA) in intensity-modulated radiotherapy (IMRT). Contoured left and right head phantoms were converted from DICOM to STL format. Fused deposition modeling (FDM) was used to construct an anthropomorphic patient-specific head phantom with a 3D printer. An established QA technique and the patient-specific head phantom were used to compare the calculated and measured doses. When the established technique was used to compare the calculated and measured doses, the gamma passing rate for γ ≤ 1 was 97.28%, while the gamma failure rate for γ > 1 was 2.72%. When the 3D-printed patient-specific head phantom was used, the gamma passing rate for γ ≤ 1 was 95.97%, and the gamma failure rate for γ > 1 was 4.03%. The 3D printed patient-specific head phantom was concluded to be highly feasible for patient-specific QA prior to complicated radiotherapy procedures such as IMRT.
ISSN:1932-6203
1932-6203
DOI:10.1371/journal.pone.0181560