Comparison of Manual and Automated Preprocedural Segmentation Tools to Predict the Annulus Plane Angulation and C-Arm Positioning for Transcatheter Aortic Valve Replacement
Preprocedural manual multi-slice-CT-segmentation tools (MSCT-ST) define the gold standard for planning transcatheter aortic valve replacement (TAVR). They are able to predict the perpendicular line of the aortic annulus (PPL) and to indicate the corresponding C-arm angulation (CAA). Fully automated...
Gespeichert in:
Veröffentlicht in: | PloS one 2016-04, Vol.11 (4), p.e0151918-e0151918 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Preprocedural manual multi-slice-CT-segmentation tools (MSCT-ST) define the gold standard for planning transcatheter aortic valve replacement (TAVR). They are able to predict the perpendicular line of the aortic annulus (PPL) and to indicate the corresponding C-arm angulation (CAA). Fully automated planning-tools and their clinical relevance have not been systematically evaluated in a real world setting so far.
The study population consists of an all-comers cohort of 160 consecutive TAVR patients with a drop out of 35 patients for technical and anatomical reasons. 125 TAVR patients underwent preprocedural analysis by manual (M-MSCT) and fully automated MSCT-ST (A-MSCT). Method-comparison was performed for 105 patients (Cohort A). In Cohort A, CAA was defined for each patient, and accordance within 10° between M-MSCT and A-MSCT was considered adequate for concept-proof (95% in LAO/RAO; 94% in CRAN/CAUD). Intraprocedural CAA was defined by repetitive angiograms without utilizing the preprocedural measurements. In Cohort B, intraprocedural CAA was established with the use of A-MSCT (20 patients). Using preprocedural A-MSCT to indicate the corresponding CAA, the levels of contrast medium (ml) and radiation exposure (cine runs) were reduced in Cohort B compared to Cohort A significantly (23.3±10.3 vs. 35.3 ±21.1 ml, p = 0.02; 1.6±0.7 vs. 2.4±1.4 cine runs; p = 0.02) and trends towards more safety in valve-positioning could be demonstrated.
A-MSCT-analysis provides precise preprocedural information on CAA for optimal visualization of the aortic annulus compared to the M-MSCT gold standard. Intraprocedural application of this information during TAVR significantly reduces the levels of contrast and radiation exposure.
ClinicalTrials.gov NCT01805739. |
---|---|
ISSN: | 1932-6203 1932-6203 |