Bioreactance Is Not Interchangeable with Thermodilution for Measuring Cardiac Output during Adult Liver Transplantation
Thermodilution technique using a pulmonary artery catheter is widely used for the assessment of cardiac output (CO) in patients undergoing liver transplantation. However, the unclearness of the risk-benefit ratio of this method has led to an interest in less invasive modalities. Thus, we evaluated w...
Gespeichert in:
Veröffentlicht in: | PloS one 2015-05, Vol.10 (5), p.e0127981-e0127981 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Thermodilution technique using a pulmonary artery catheter is widely used for the assessment of cardiac output (CO) in patients undergoing liver transplantation. However, the unclearness of the risk-benefit ratio of this method has led to an interest in less invasive modalities. Thus, we evaluated whether noninvasive bioreactance CO monitoring is interchangeable with thermodilution technique.
Nineteen recipients undergoing adult-to-adult living donor liver transplantation were enrolled in this prospective observational study. COs were recorded automatically by the two devices and compared simultaneously at 3-minute intervals. The Bland-Altman plot was used to evaluate the agreement between bioreactance and thermodilution. Clinically acceptable agreement was defined as a percentage error of limits of agreement 92%.
A total of 2640 datasets were collected. The mean CO difference between the two techniques was 0.9 l/min, and the 95% limits of agreement were -3.5 l/min and 5.4 l/min with a percentage error of 53.9%. The percentage errors in the dissection, anhepatic, and reperfusion phase were 50.6%, 56.1%, and 53.5%, respectively. The concordance rate between the two techniques was 54.8%.
Bioreactance and thermodilution failed to show acceptable interchangeability in terms of both estimating CO and tracking CO changes in patients undergoing liver transplantation. Thus, the use of bioreactance as an alternative CO monitoring to thermodilution, in spite of its noninvasiveness, would be hard to recommend in these surgical patients. |
---|---|
ISSN: | 1932-6203 1932-6203 |
DOI: | 10.1371/journal.pone.0127981 |