A recommendation for revised dose calibrator measurement procedures for 89Zr and 124I

Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as sim...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:PloS one 2014-09, Vol.9 (9), p.e106868-e106868
Hauptverfasser: Beattie, Bradley J, Pentlow, Keith S, O'Donoghue, Joseph, Humm, John L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89)Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R) using a published calibration setting number of 465. PET images of (124)I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89)Zr and (124)I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an 18F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89)Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124)I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89)Zr and (124)I. These and other positron emitting radionuclides can be accurately calibrated relative to 18F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances.
ISSN:1932-6203
1932-6203
DOI:10.1371/journal.pone.0106868