Analysis of four aberrometers for evaluating lower and higher order aberrations
To compare the measurements of lower and higher order aberrations (HOA) of 4 commonly used aberrometers. Massachusetts Eye & Ear Infirmary, Boston, USA. Prospective, cross-sectional study, in a controlled, single-blinded fashion. Multiple readings were obtained in 42 eyes of 21 healthy volunteer...
Gespeichert in:
Veröffentlicht in: | PloS one 2013-01, Vol.8 (1), p.e54990-e54990 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To compare the measurements of lower and higher order aberrations (HOA) of 4 commonly used aberrometers.
Massachusetts Eye & Ear Infirmary, Boston, USA.
Prospective, cross-sectional study, in a controlled, single-blinded fashion.
Multiple readings were obtained in 42 eyes of 21 healthy volunteers, at a single visit, with each of the following aberrometers: Alcon LADARWave®, Visx WaveScan®, B & L Zywave®, and Wavelight Allegro Analyzer®. Results were compared and analyzed in regards to the lower and HOA, to the different wavefront sensing devices and software, Tscherning and Hartmann-Shack and between the Fourier and Zernike algorithms. Statistical analysis included Bland-Altman plots, Intraclass Correlation Coefficient (ICC), multiple comparison tests with Analysis of Variance and Kruskal-Wallis. Significant level was set to p |
---|---|
ISSN: | 1932-6203 1932-6203 |
DOI: | 10.1371/journal.pone.0054990 |