Evaluation of portable point-of-care CD4 counter with high sensitivity for detecting patients eligible for antiretroviral therapy
Accurate, inexpensive point-of-care CD4+ T cell testing technologies are needed that can deliver CD4+ T cell results at lower level health centers or community outreach voluntary counseling and testing. We sought to evaluate a point-of-care CD4+ T cell counter, the Pima CD4 Test System, a portable,...
Gespeichert in:
Veröffentlicht in: | PloS one 2012-04, Vol.7 (4), p.e34319-e34319 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurate, inexpensive point-of-care CD4+ T cell testing technologies are needed that can deliver CD4+ T cell results at lower level health centers or community outreach voluntary counseling and testing. We sought to evaluate a point-of-care CD4+ T cell counter, the Pima CD4 Test System, a portable, battery-operated bench-top instrument that is designed to use finger stick blood samples suitable for field use in conjunction with rapid HIV testing.
Duplicate measurements were performed on both capillary and venous samples using Pima CD4 analyzers, compared to the BD FACSCalibur (reference method). The mean bias was estimated by paired Student's t-test. Bland Altman plots were used to assess agreement.
206 participants were enrolled with a median CD4 count of 396 (range; 18-1500). The finger stick PIMA had a mean bias of -66.3 cells/µL (95%CI -83.4-49.2, P500 cells/µL) with a mean bias of -120.6 (95%CI -162.8, -78.4, P |
---|---|