Pulsed NMR probe‐monitoring system
A simple device for measuring the instantaneous reflected power from a NMR probe is described. It is built from commercially available directional power‐meter components coupled to a low‐speed oscilloscope. The device can be used to monitor the performance of a probe while data are being acquired. I...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 1987-06, Vol.58 (6), p.1128-1130 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple device for measuring the instantaneous reflected power from a NMR probe is described. It is built from commercially available directional power‐meter components coupled to a low‐speed oscilloscope. The device can be used to monitor the performance of a probe while data are being acquired. It is particularly useful for adjusting probes whose characteristics change during a high‐power pulse because of component heating. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1139571 |