Manufacture, observation, and test of membranes with locatable single pores

A method for generating single pores down to 0.1 μm diameter in the center of a large circular foil is described, based on nuclear tracks. The foil is framed by a tension ring which enables one to handle the foils in a well‐defined precise way. The single pore has a lateral displacement of ±0.1 mm w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Rev. Sci. Instrum.; (United States) 1986-08, Vol.57 (8), p.1654-1660
Hauptverfasser: Packard, R. E., Pekola, J. P., Price, P. B., Spohr, R. N. R., Westmacott, K. H., Yu‐Qun, Zhu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for generating single pores down to 0.1 μm diameter in the center of a large circular foil is described, based on nuclear tracks. The foil is framed by a tension ring which enables one to handle the foils in a well‐defined precise way. The single pore has a lateral displacement of ±0.1 mm with respect to the tension ring center. The foils used are polycarbonate of the type Makrofol and have thicknesses between 2 and 10 μm. For calibration of the single pore diameters, multiple nuclear tracks between 0.1 and 3.5 μm diameter are etched and observed by microscopy. The microscopic observations are compared with gas‐flow measurements, using two alternative methods: multiple holes are tested under viscous flow conditions of N2 gas at normal temperature and pressure; single holes are tested under collisionless flow conditions of 4He gas at liquid‐nitrogen temperature, using a capacitance method.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1138545