Phase-aberration correction using signals from point reflectors and diffuse scatterers: measurements
A method for phase-aberration correction of phased-array images is tested using a model of near-field velocity inhomogeneities. A set of grooved room-temperature vulcanizing plates was constructed to simulate near-field aberrations encountered in clinical ultrasound imaging. As expected, large image...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 1988-11, Vol.35 (6), p.768-774 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A method for phase-aberration correction of phased-array images is tested using a model of near-field velocity inhomogeneities. A set of grooved room-temperature vulcanizing plates was constructed to simulate near-field aberrations encountered in clinical ultrasound imaging. As expected, large image distortion was experienced when grooved plates producing significant aberrations were placed near the surface of the array. An iterative aberration correction procedure based on cross-correlation measures between neighboring elements in a phased array, using signals reflected from diffuse scatterers, significantly reduced the effects of these aberrations, producing images nearly identical to those generated in the absence of aberrations. The results suggest that a practical phase-aberration correction system can be constructed for medical ultrasound imaging and possibly all coherent imaging systems by using a sampled aperture.< > |
---|---|
ISSN: | 0885-3010 1525-8955 |
DOI: | 10.1109/58.9334 |