Hard x‐ray diagnostics for laser fusion plasmas analysis

Two diagnostics for the study of hard x‐ray emission produced by plasmas during inertial confinement fusion experiments, are described: (i) a hard x‐ray spectrometer to measure the high‐energy tail of the spectrum, with eight spectral bands in the 50–400‐keV range, by fluorescence and simple filter...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of Scientific Instruments 1992-10, Vol.63 (10), p.5097-5100
Hauptverfasser: Miquel, J‐L., Bourgade, J‐L., Schirmann, D., Veaux, J., Gex, J‐P., Cavailler, C., de Mascureau, J., Frotté, V., Trochet, R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two diagnostics for the study of hard x‐ray emission produced by plasmas during inertial confinement fusion experiments, are described: (i) a hard x‐ray spectrometer to measure the high‐energy tail of the spectrum, with eight spectral bands in the 50–400‐keV range, by fluorescence and simple filter techniques; (ii) a time‐integrated pinhole imaging device using a microchannel plate converter tube and a 7.5‐μm‐thick tantalum photocathode adjusted for a mean spectral sensitivity around 200 keV. Adjustment and test of these diagnostics have been performed with a pulsed hard x‐ray generator using a plasma focus device. Absolute calibration of the spectrometer detectors (scintillator Photomultiplier) was obtained on an x‐ray tube with fluorescent emission lines (10–100 keV) and with a cobalt radioactive source at 1.23 MeV.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1143450