Digital impedance bridge

An impedance bridge that compares two-terminal standard inductors to characterized AC resistors in the frequency range of 10 Hz to 100 kHz is described. A dual-channel, digitally-synthesized source and sampling digital multimeter are used to generate and measure relevant bridge signals. A linear int...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 1995-04, Vol.44 (2), p.436-439
Hauptverfasser: Waltrip, B.C., Oldham, N.M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An impedance bridge that compares two-terminal standard inductors to characterized AC resistors in the frequency range of 10 Hz to 100 kHz is described. A dual-channel, digitally-synthesized source and sampling digital multimeter are used to generate and measure relevant bridge signals. A linear interpolation algorithm is used to autocalibrate the bridge to a 1 nF gas dielectric capacitor. An intercomparison of the new bridge with existing measurement standards conducted in the low audio frequency range shows agreement of 50 to 200 parts in 10/sup 6/ for inductors for 1 mH to 10 H.< >
ISSN:0018-9456
1557-9662