Digital impedance bridge
An impedance bridge that compares two-terminal standard inductors to characterized AC resistors in the frequency range of 10 Hz to 100 kHz is described. A dual-channel, digitally-synthesized source and sampling digital multimeter are used to generate and measure relevant bridge signals. A linear int...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 1995-04, Vol.44 (2), p.436-439 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An impedance bridge that compares two-terminal standard inductors to characterized AC resistors in the frequency range of 10 Hz to 100 kHz is described. A dual-channel, digitally-synthesized source and sampling digital multimeter are used to generate and measure relevant bridge signals. A linear interpolation algorithm is used to autocalibrate the bridge to a 1 nF gas dielectric capacitor. An intercomparison of the new bridge with existing measurement standards conducted in the low audio frequency range shows agreement of 50 to 200 parts in 10/sup 6/ for inductors for 1 mH to 10 H.< > |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/19.377873 |